![Nellcor OXIMAX N-560 Скачать руководство пользователя страница 23](http://html2.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_service-manual_3907265023.webp)
Performance Verification
N-560 Service Manual
13
Figure 5: Adjusting Low %SpO
2
Alarm Limit
8. Press the
SpO
2
Alarm Limit
button two times rapidly. Within three
seconds, press and hold the
Adjust Down
button and verify that the
%SpO
2
display reduces to a minimum of “20.” Press and hold the
Adjust
Up
button and verify that the %SpO
2
display cannot be raised past the
upper alarm limit setting minus one point.
9. Turn the N-560 off by pressing the
Power On/Off
button.
10. Turn the N-560 on by pressing the
Power On/Off
button.
11. Wait for POST to be completed. The %SpO
2
and pulse rate alarm limits
will be reset to the default values.
12. Press the
Pulse Rate Alarm Limit
button. Verify that the N-560 emits a
single beep, the
Pulse Rate
display indicates an alarm limit of “170”
and that the %SpO
2
display window shows “HI” for about 3 seconds.
Figure 6: Adjusting High Pulse Rate Alarm Limit
13. Press the
Pulse Rate Alarm Limit
Button. Within 3 seconds press and
hold the
Adjust Down
button. Verify that the
Pulse Rate
display
reduces to a minimum of one point above the pulse rate low limit alarm
setting.
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......