Performance Verification
10
3. The N-560 automatically starts the Power-On Self-Test (POST), which
tests N-560 circuitry and functions.
CAUTION: During POST (immediately after power-up), confirm that all
display segments and indicators light, and the speaker sounds a
one-second pass tone.
4. While performing POST, the self-test display appears for approximately
five seconds. During this time:
• All indicators illuminate
• All segments of all numeric digits light and change from red to green
• All segments of the Pulse Amplitude Display light
• All segments of the
SatSeconds
indicator light
5. Once the display test portion of POST is complete, the N-560 version is
displayed for approximately two seconds.