EasyOne Pro™ / EasyOne Pro™ LAB
81
Performing a Test
Note
The parameter names used by ndd Medizintechnik AG are
those standardized by ATS and ERS. (See literature
references [14] and[11] (
For forced spirometric tests and depending on your configuration, the
"Best" column in the result table shows the "Best Value" of all accepted tri-
als or the "Best Trial", i.e., the value of the best trial (
The most important parameters will be explained below:
For slow spirometry (SVC), the mean value of the accepted trials will be
shown with the following exceptions: VT, ERV, IRV, IC.
For DLCO, the mean values will be displayed in the "Result" column for the
following parameters: DLCO, VA, TLCO, DLAdj and TLC.
IC
Inspiratory Capacity
inspiratory capacity
(VT + IRV)
MVV
Maximum Voluntary Venti-
lation
maximum ventilation vol-
ume per minute
LLN
Lower Limit of Normal
lower limit of normal
Lung Age
lung age, see section
DLCO
CO diffusing capacity
(ml/min/mmHg)
CO diffusing capacity
(ml/min/mmHg)
TLCO
CO diffusing capacity
(mmol/min/kPa)
CO diffusing capacity
(mmol/min/kPa)
DLadj
Diffusing capacity
adjusted (Hb, PaO
2
, COhb,
altitude)
adjusted diffusing capac-
ity
VA
Alveolar volume (BTPS)
alveolar volume (BTPS)
VI
Inspiratory Capacity from
DLCO maneuver
inspiratory capacity from a
DLCO test
DLCO/VA
Krogh Factor
Krogh factor
TLC
Total Lung Capacity
total lung capacity