EasyOne Pro™ / EasyOne Pro™ LAB
51
Performing a Test
Adding a Spirometry Trial
If you would like to add trials to a previous test, e.g. if the patient needed a
break or if other patients were tested in between, please proceed as fol-
lows. Remember, however, that it is only possible to add a trial to a previous
test that was performed on the same day.
Select the patient.
Click
Perform Test
.
Select the test, e.g.
FVC
.
A menu will appear.
Click
Add Trial
or
Add
Post
.
The acquisition screen will be displayed.
Perform the new spirometry maneuvers.
FVL Test (inspiratory and expiratory measurement)
Ambient Conditions
The ATPS values (
A
mbient
T
emperature
P
ressure
S
aturated = spirometer
conditions) of inspiratory measurements must be converted to BTPS values
(
B
ody
T
emperature
P
ressure
S
aturated = body conditions). For this conver-
sion, the system needs the following data:
relative humidity
room temperature
atmospheric pressure or altitude.
Before the inspiratory test can be started, a window pops up with these
data. Check the data. For the pending test you can now edit the data, if nec-
essary, or confirm them, if they are correct.