Specifications
132
EasyOne Pro™ / EasyOne Pro™ LAB
Predicted val-
ues (spirome-
try)
NHANES III, Knudson_83, Knudson_76, Crapo, Mor-
ris, Hsu, Dockery (Harvard), Polgar, Gutierrez
(Canada), Eigen, Pereira, Pereira 2006/2008, Pérez-
Padilla (PLATINO), Pérez Padilla (Mexico), Pérez
Padilla (Mexico, Ped.), Chile 2010, 1997, ERS (ECCS/
EGKS), Zapletal 1977, Zapletal 2003, Stanojevic 2009
(GLI), Quanjer 2012 (GLI), Rosenthal, Austria 1988,
Austria 1994, Sapaldia (Swiss), Roca (Spain, SEPAR),
Vilozni 2005, Falaschetti, Klement (Russia), Heden-
ström (Sweden), Gulsvik, (Norway), Berglund, Birath
(Sweden), Langhammer (Norway), Finnish, Nystad,
Hibbert, Gore, Crockett, Ethiopia, Asia 1-4, JRS 2001,
Fukuda Standard (see
Predicted val-
ues (lung
capacity and
DLCO)
Ayers, Burrows, Crapo, Goldman & Becklake, Knud-
son, McGrath & Thompson, Miller, Gutierrez (Canada),
Klement (Russia), NHANES, Polgar, ERS, Zapletal,
Roca, Hedenström, Gulsvik, Pereira 2008, Thompson
Operating
conditions
Temperature
5 to 40 °C (41 to 104 °F)
LAB: 10 to 40 °C
(50 to 104 °F)
Relative humidity
15% to 95%,
no condensation
LAB: 30% to 75%,
no condensation
Atmospheric pressure
700 to 1060 hPa
Storage/
transport
Temperature
-20 to 50 °C (-4 to +122 °F)
Relative humidity
5% to 95%, no
condensation
Atmospheric pressure
500 to 1060 hPa
Certifications
CE Declaration of conformity
Device classi-
fication
Protection class I
Type BF applied part
Furthermore the device is not intended to be used in
the presence of oxygen rich environments. However,
where the device contains oxygen rich environments,
it complies with applicable requirements for fire pre-
vention.
Operating
voltage range
100 to 240 V, 50 to 60 Hz