EasyOne Pro™ / EasyOne Pro™ LAB
131
Specifications
10 Specifications
Dimensions
270 x 335 x 270 mm
3
(h x w x d)
Weight [kg]
< 8 kg, LAB < 9 kg
Flow/volume
measurement
Accuracy, volume:
±2% or 0.050 l
Accuracy, flow:
±2% or 0.020 l/s
(except PEF)
Accuracy, PEF:
±5% or 0.200 l
Accuracy, MVV:
±5% or 5 l/min
Resolution, volume
>1 ml
Resolution, flow
>4 ml/s
Measuring range, volume ±12 l
Measuring range, flow
±16 l/s
Resistance
<1.5 cm H
2
O/l/s at 12 l/s
Molar mass
analyzer
Type
ultrasonic transit time
Measuring range
9 to 50 g/mol
Accuracy
<0.02 g/mol
Resolution
0.01 g/mol
Carbon mon-
oxide analyzer
Type NDIR
Measuring range
0 to 0.35%
Accuracy
±0.001%
Resolution
0.0001%
CO2 analyzer
(LAB only)
Type nondispersive
infrared
Measuring range
0 to 15%
Resolution
0.005%
Accuracy
from 0 to 5%:±0.05%
from 5 to 10%:±0.20%
>10%: not specified
Mouth pres-
sure trans-
ducer
Measuring range
±100 mbar
Accuracy
±5%