![National Instruments CompactDAQ cDAQ-9185 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/national-instruments/compactdaq-cdaq-9185/compactdaq-cdaq-9185_user-manual_3403867021.webp)
1-10
|
ni.com
Chapter 1
Getting Started with the cDAQ Chassis
power to all of the modules. The green POWER LED on the front panel identifies when the
power input is in use.
Complete the following steps to connect a power source to the cDAQ chassis.
1.
Make sure the power source is turned off.
2.
If connected, loosen the connector screw flanges and remove the power screw terminal
connector plug from the cDAQ chassis. Figure 1-7 shows the terminal screws, which secure
the wires in the screw terminals, and the connector screw flanges, which secure the
connector plug on the front panel.
Figure 1-7.
Power Screw Terminal Connector Plug
Caution
Do
not
tighten or loosen the terminal screws on the power connector while
the power is on.
3.
Connect the positive lead of the power source to the V terminal of the power connector plug
and tighten the terminal screw.
4.
Connect the negative lead of the power source to the C terminal of the power screw
terminal connector plug and tighten the terminal screw.
5.
Install the power connector plug on the front panel of the cDAQ chassis and tighten the
connector screw flanges.
6.
Turn on the external power source.
If the power source is connected to the power connector using long wiring with high DC
resistance, the voltage at the power connector may be significantly lower than the specified
voltage of the power source.
The C terminal is not connected to chassis ground. You can connect the C terminal to chassis
ground externally. Refer to the
Power Requirements
section of the specifications document for
your chassis for information about the power supply input range. Refer to the
Safety Voltages
section of the specifications document for your chassis
for information about the maximum
voltage from terminal to chassis ground.
1
V (Positive) Terminal Screw
2
C (Negative) Terminal Screw
3
Connector Screw Flanges
1
2
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com