6-37
cDAQ-9185/9189 User Manual
Some of these options may not be available in some driver software. Refer to the
Device Routing
in MAX
topic in the
NI-DAQmx Help
or the
LabVIEW Help
for more information about available
routing options.
Counter
n
Sample Clock Signal
Use the Counter
n
Sample Clock (Ctr
n
SampleClock) signal to perform sample clocked
acquisitions and generations.
You can specify an internal or external source for Counter
n
Sample Clock. You also can specify
whether the measurement sample begins on the rising edge or falling edge of Counter
n
Sample
Clock.
If the cDAQ chassis receives a Counter
n
Sample Clock when the FIFO is full, it reports an
overflow error to the host software.
Using an Internal Source
To use Counter
n
Sample Clock with an internal source, specify the signal source and the
polarity of the signal. The source can be any of the following signals:
•
DI Sample Clock
•
DO Sample Clock
•
AI Sample Clock (ai/SampleClock, te0/SampleClock, te1/SampleClock)
•
AI Convert Clock
•
AO Sample Clock
•
DI Change Detection output
Several other internal signals can be routed to Counter
n
Sample Clock through internal routes.
Refer to
Device Routing in MAX
in the
NI-DAQmx Help
or the
LabVIEW Help
for more
information.
Using an External Source
You can route any of the following signals as Counter
n
Sample Clock:
•
Any PFI terminal
•
Analog Comparison Event
You can sample data on the rising or falling edge of Counter
n
Sample Clock.
Routing Counter
n
Sample Clock to an Output Terminal
You can route Counter
n
Sample Clock out to any PFI terminal. The PFI circuitry inverts the
polarity of Counter
n
Sample Clock before driving the PFI terminal.