cDAQ-9185/9189 User Manual
Routing a Signal to Counter
n
Gate
Each counter has independent input selectors for the Counter
n
Gate signal. Any of the following
signals can be routed to the Counter
n
Gate input:
•
Any PFI terminal
•
AI Reference Trigger
•
AI Start Trigger
•
AO Sample Clock
•
DI Sample Clock
•
DI Reference Trigger
•
DO Sample Clock
•
Change Detection Event
•
Analog Comparison Event
In addition, a counter’s Internal Output or Source can be routed to a different counter’s gate.
Some of these options may not be available in some driver software. Refer to the
Device Routing
in MAX
topic in the
NI-DAQmx Help
or the
LabVIEW Help
for more information about available
routing options.
Routing Counter
n
Gate to an Output Terminal
You can route Counter
n
Gate out to any PFI terminal.
Counter
n
Aux Signal
The Counter
n
Aux signal indicates the first edge in a two-signal edge-separation measurement.
Routing a Signal to Counter
n
Aux
Each counter has independent input selectors for the Counter
n
Aux signal. Any of the following
signals can be routed to the Counter
n
Aux input:
•
Any PFI terminal
•
AI Reference Trigger
•
AI Start Trigger
•
Analog Comparison Event
•
Change Detection Event
In addition, a counter’s Internal Output, Gate or Source can be routed to a different counter’s
Aux. A counter’s own gate can also be routed to its Aux input.
Some of these options may not be available in some driver software. Refer to the
Device Routing
in MAX
topic in the
NI-DAQmx Help
or the
LabVIEW Help
for more information about available
routing options.