Chapter 2
Configuration and Installation
© National Instruments Corporation
2-31
AT-MIO-64F-5 User Manual
The pulse width is defined as 500 nsec minimum. The EXTSTROBE* signal can be used by an
external device to latch signals or trigger events. The EXTSTROBE* signal is an HCT signal.
EXTCONV* Signal
A/D conversions can be externally triggered with the EXTCONV* pin. Applying an active low
pulse to the EXTCONV* signal initiates an A/D conversion. Figure 2-14 shows the timing
requirements for the EXTCONV* signal.
tw 50 nsec minimum
ADC switches to hold mode within 100 nsec from this point
V
IL
VIH
tw
tw
Figure 2-14. EXTCONV* Signal Timing
The minimum allowed pulse width is 50 nsec. The ADC switches to hold mode within 100 nsec
of the high-to-low edge. This hold mode delay time is a function of temperature and does not
vary from one conversion to the next. There is no maximum pulse width limitation.
EXTCONV* should be high for at least one conversion period before going low. The
EXTCONV* signal is one HCT load and is pulled up to +5 V through a 10 k
W
resistor.
EXTCONV* is also driven by the output of Counter 3 of the Am9513A Counter/Timer. This
counter is also referred to as the sample-interval counter. The output of Counter 3 and the RTSI
connection to EXTCONV* must be disabled to a high-impedance state if A/D conversions are to
be controlled by pulses applied to the EXTCONV* pin. If Counter 3 is used to control A/D
conversions, its output signal can be monitored at the EXTCONV* pin.
A/D conversions generated by either the EXTCONV* signal or the sample-interval counter are
inhibited outside of a data acquisition sequence and when gated by either the hardware
(EXTGATE*) signal or software command register gate.
Note:
EXTCONV* and the output of Counter 3 of the Am9513A are physically connected
together on the AT-MIO-64F-5. If Counter 3 is used in an application, the EXTCONV*
signal must be left undriven. Conversely, if EXTCONV* is used in an application,
Counter 3 must be disabled.
EXTTRIG* Signal
Any data acquisition sequence can be initiated by an external trigger applied to the EXTTRIG*
pin. Applying a falling edge to the EXTTRIG* pin starts the sample and sample-interval
counters, thereby initiating a data acquisition sequence. Figure 2-15 shows the timing
requirements for the EXTTRIG* signal.
Содержание AT-MIO-64F-5
Страница 13: ......
Страница 16: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......
Страница 219: ......
Страница 220: ......
Страница 221: ......
Страница 222: ......
Страница 223: ......
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 227: ......
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......