7-18 

  Remote control 

 

 

 

7.6.3 Analyzer  
           Reply 
 

The PMM 7010 replies to the command “

SAGO

” sending back an array of bytes 

which contains all the information needed to draw a sweep. 
Typically, the user should first send the PMM 7010 all the setting commands to 
insure the receiver is correctly set on the wanted parameters and then read the 
reply. 
Just as an example, the user could send the following commands: 

 

7.6.3.1 Reply  
         example 
 

#SAAT -1* 

Set Attenuator Automatic mode 

#SAPA Off* 

Switch Preamplifier OFF 

#SPLM a* 

Switch Pulse Limiter OFF 

#SAFF 
150e3,30e6* 

Sets the Start frequency to 150kHz 
and Stop frequency to 30MHz 

#SAHT 2* 

Sets the hold time to 2 ms 

#SADT 2* 

Sets the detector to AVG 

............. More 

Command 

............. More 

Command 

#SAGO * 

Start Sweep in Analyzer Mode 

 

 

 
The PMM7010 replies by sending a 48 bytes header followed by a variable 
size array of integer (the size depends on frequency settings) as follows: 
 

 

HEADER 

ID Byte   Value 

Meaning 

The first 8 Bytes are just a fixed header and acknowledge the command 
 

2 G 
3 O 
4 = 
5 O 
6 K 
7 0x13 

(CR)   

8 0x10 

(LF)   

9 Start 

Frequency 

Little Endian 32 bit Float figure representing the start frequency of sweep 

10 
11 
12 
13 Stop 

Frequency 

Little Endian 32 bit Float figure representing the Stop frequency of sweep 

14 
15 
16 
17 Step 

Frequency 

Little Endian 32 bit Float figure representing the step frequency of sweep 

18 
19 
20 
21 
… 
26 

Reserved 
… 

 
… 

27 

Attenuator 

Little Endian 16 bit integer representing the attenuator used 

28 
29 
… 
48 

Reserved 
…. 

 
… 

Содержание PMM 7010 Series

Страница 1: ...e first four digits and the letter are the Serial Number prefix the last five digits are the Serial Number suffix The prefix is the same for identical instruments it changes only when a configuration...

Страница 2: ...TY SYMBOLS You now own a high quality instrument that will give you many years of reliable service Nevertheless even this product will eventually become obsolete When that time comes please remember t...

Страница 3: ...h LINE led 2 6 Environment 2 7 Return for Service 2 8 Equipment Cleaning 2 9 Equipment ventilation 2 10 Hardware Installation 2 11 Using an Artificial Mains Network AMN or LISN 2 12 Using Pulse Limite...

Страница 4: ...r Rack 19 6 1 Introduction 6 2 Instruments Items 6 3 Optional accessories 6 4 PMM 9010 RMA Main Specifications 6 5 PMM 9010 RMA Front view 6 6 PMM 9010 RMA Inside view 6 7 Rack requirements 6 8 Requir...

Страница 5: ...M LISN three phase remote cable configuration for 7010 PMM L3 25 remote cable configuration for PMM 7010 AMN Principle a type or T type LISN b V type LISN Example of test Setup for RFI Voltage Measure...

Страница 6: ...be provided with fuses or protections rated to the unit Verify that the unit is set to match the available mains voltage and correct fuse rating is installed before applying power The Safety Class I...

Страница 7: ...base delle seguenti norme europee armonizzate applicate con esito positivo based on the following harmonized European Standards successfully applied EMC Emissioni EMC Emission EN 61326 1 2013 EMC Imm...

Страница 8: ...VIII Contents This page has been left blank intentionally...

Страница 9: ...to 1000 MHz Option 7010 00 9 kHz up to 1000 MHz Option 7010 01 9 kHz up to 30 MHz Option 7010 02 and 9 kHz up to 3000 MHz Option 7010 03 All measurements performed by the PMM 7010 are according to the...

Страница 10: ...Switching Box for LISNs and Loop Antennas LISN Service Kit AC BNC adapter for LISNs verification and calibration SHC 1 1000 35 dB CISPR Voltage probe 1500 SHC 2 1000 30 dB CISPR Voltage probe 1500 RA...

Страница 11: ...50 N fem 1 2 2 over 1 GHz 1 2 1 4 over 30 MHz 2 over 1 GHz 0 dB to 35 dB 5dB steps 0 dB to 50 dB 2dB steps over 30 MHz Built in up to 30 MHz Max input level Sinewave AC voltage Pulse spectral density...

Страница 12: ...dB A m dBpT Measurement accuracy S N 20 dB 9 kHz to 1 GHz 1 0 dB 1 GHz to 3 GHz 1 5 dB CISPR 16 1 1 conformity Standard compliant QP detector down to 20 Hz PRF Demodulation AM FM Internal loudspeaker...

Страница 13: ...es if the signal source is from the internal LISN or from the N connector L1 L2 These two leds indicate the line of the internal LISN under investigation Artificial Mains Network 16A It is the mains s...

Страница 14: ...d Serial Number Mains socket to connect the EUT to the mains through the internal LISN Earth ground connector USER PORT User I O Port Fan Cooling Fan controlled by firmware RS232 9 pin DB9 connector O...

Страница 15: ...n the CISPR bands A 9 kHz 150 kHz B 150 kHz 30 MHz C 30 300 MHz and D 300 1000 MHz the standards requires the use of specially shaped 200 Hz 9 kHz and 120 kHz filters Thanks to its architecture and to...

Страница 16: ...mple and reliable to use to be the base building block for any possible emission system to measure and evaluate any electric or electronic device from the very first design stages to the final certifi...

Страница 17: ...l as well as to your NARDA Representative To avoid further damage do not turn on the instrument when there are signs of shipping damage to any portion of it 2 4 Preparation for Use This is a Safety Cl...

Страница 18: ...re only occurs under certain conditions explain how to duplicate the failure If possible reusing of the original packaging to ship the equipment is preferable In case other package should be used ensu...

Страница 19: ...ly The boot sequence is very fast and the receiver is ready to use in a few seconds after having switched the ON button When the initial sequence is completed the receiver is ready to be controlled by...

Страница 20: ...Equipment Under Test supply ON or OFF 2 13 Using Current and Voltage Probes When a LISN cannot be used e g when measurements have to be made on terminals other than the mains ones such as load or com...

Страница 21: ...be programmed and managed the connector has the following hardware connection PIN Signal 1 nc 2 nc 3 12 VDC max 50 mA 4 OUT 0 5 OUT 2 6 nc 7 nc 8 and 9 nc 10 IN 0 11 GND 12 OUT 1 13 nc 14 nc 15 nc Da...

Страница 22: ...oop Antenna Remote Cable configuration for PMM 7010 The following figure shows the HXYZ 9170 Triple Loop Antenna remote cable pin configuration This cable can be requested to Narda or arranged locally...

Страница 23: ...emote Cable configuration for PMM 7010 L3 25 excluded The following figure shows the LISN remote cable pin configuration The cable can be requested to Narda or arranged locally Fig 2 3 PMM LISN three...

Страница 24: ...L3 25 Remote Cable configuration for PMM 7010 The following figure shows the LISN remote cable pin configuration The cable can be requested to Narda or arranged locally Fig 2 4 PMM L3 25 remote cable...

Страница 25: ...r the control of the PMM Emission Suite PC Software it is possible to use the receiver in different modes of operation 3 2 1 SWEEP MODE The Sweep Mode is the most common way to perform measurements in...

Страница 26: ...n the main menu and immediately the first Manual window opens In manual mode the levels corresponding to the detectors in use are displayed both in analogue and in digital format and exactly as for th...

Страница 27: ...s in accordance with the following rule The Amplitude is proportional to repetition frequency 1 2 The value is the maximum level during the measurement time Hold Time and the averaging is carried out...

Страница 28: ...emember to enable a limit to run the Smart detector function Selecting one out of the five Smart detectors the receiver will execute the scan using at first the Peak detector and if a peak value is fo...

Страница 29: ...time starts again 3 4 RBW Filters The Resolution Bandwidth box is used to select the bandwidth of the measuring filter The drop down menu allows the User to select from the list the filter desired Th...

Страница 30: ...nal is gone so the receiver lowers the attenuation but then a new peak arrives and so on and so forth On the other hand if the Hold Time is too high the PMM 7010 cannot properly follow the signals In...

Страница 31: ...rd and return lines Asymmetrical or common mode voltages These EMI components are measured between the electrical midpoint of the two conductor voltages and reference ground Usually these are the comp...

Страница 32: ...and data lines Standard impedance is 150 as well An Artificial Mains Network shall be designed in order to 1 terminate each line power signal etc of the EUT Equipment Under Test with a standardized i...

Страница 33: ...has a standard impedance Zin 1 5 K 10 pF Voltage Probe is used for measuring the unsymmetrical RFI voltage when it is not possible to carry out measurement by interconnecting an Artificial V type Netw...

Страница 34: ...B V press Esc 3 Touch the Measure key then Freq B Band 4 Then Level Input and set the Minimum Attenuation to 10 dB and the Attenuation is automatic press Esc 5 The Pulse Limiter is always enabled to p...

Страница 35: ...low ambient interferences are often sufficient for a preliminary evaluation Conducted measurements do not strictly require any anechoic environment while radiated tests may need it if so specified by...

Страница 36: ...4 6 Applications This page has been left blank intentionally...

Страница 37: ...a free RS232 port or USB in case of Adapter s use on the PC side The first free RS 232 port will be automatically detected by the Firmware Update Program during installation In case of troubles please...

Страница 38: ...ure the power to the PMM 7010 is stable and reliable before performing the FW Upgrade otherwise the upgrade progress could not terminate properly Anyway even in case of failure the internal BIOS will...

Страница 39: ...the process simply switch PMM 7010 on select Update Firmware or Update FPGA button and wait until the automatic transfer is completed During the firmware storing procedure a blue bar will progress fr...

Страница 40: ...ailure an error message is showed instead After Firmware Update is successfully completed switch OFF and then ON again the PMM 7010 looking at screen and then perform Auto calibration It is now possib...

Страница 41: ...urn on the PMM 7010 and connect it to a free USB or RS232 port of the PC Browse for All Programs from the Start Menu and reach the WrDongle executable link Main window displayed after the program WrDo...

Страница 42: ...nd press the button below related to the specific Option In case of failure an error message is showed instead Always cycle OFF and ON the receiver to properly initialize the function With the PMM 701...

Страница 43: ...s 482 6 mm wide including edges that protrude on each side which allow the module to be fastened to the rack frame with screws 6 2 Instrument Items PMM 9010 RMA Rack Mount Adapter includes the followi...

Страница 44: ...e 0 to 40 C Rack unit 3U Dimensions 483 x 410 x 132 5 mm WxDxH Weight 5 5 Kg without instruments 6 5 PMM 9010 RMA Front view Fig 6 1 Front view Legend 1 Holes Flanges used to fix the Rack Mount to the...

Страница 45: ...side view Fig 6 2 Inside view Legend 1 Round holes of service 2 Rectangular holes used to put 9010 9010 03P 30P 60P or 7010 Instrument 3 9010 9010 03P 30P 60P or 7010 straps 4 Rectangular holes for fu...

Страница 46: ...ave the following items Crosshead screwdriver not provided Flat blade screwdriver or rack insertion tool not provided Tape measure and level not provided Rack Mount kit provided 6 9 Moving chassis Whe...

Страница 47: ...nsure that there is adequate space around the rack to allow for servicing the chassis and for airflow If the rack has wheels ensure that the brakes are engaged or that the rack is otherwise stabilized...

Страница 48: ...ertical rails with respect to the Rack Unit Boundary and the distance show below You can use a rack insertion tool or a flat blade screwdriver to install the cage nuts Insert the chassis into the rack...

Страница 49: ...lled knob without loosing it completely Using the handles open the PMM 9010 RMA slowly Place the instrument by the rectangular holes and drag the instrument in contact with the frontal panel To ensure...

Страница 50: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...

Страница 51: ...ol identify the type of communication used The communication is at 115200 bit sec with 8 bit words one start bit one stop bit and no parity 115200 N 8 1 7 3 Protocol Be aware that only the PC can send...

Страница 52: ...e DET Reports the five detector for manual mode DMD Requests the status of demodulator DMV Requests the volume of demodulator FPGA Sends back a string containing information about version of FPGA IDN...

Страница 53: ...Creates a conversion factor frequency by frequency SDMD a Switches the Demodulator On or OFF SDMV v Sets the volume of Demodulator SLDW n freq levq leva Sets the custom double value limit frequency b...

Страница 54: ...uency in exponential notation for analyzer mode Unit is fixed MHz Example of reply ACE 1 500000e 07 which means that the Center frequency in Analyzer mode is 15MHz ADT This query command ADT sends bac...

Страница 55: ...ditional string OVER is appended at the end whenever an over range condition occurred Example of reply DET 23 22 17 09 16 23 11 36 16 01 15 50 which means Peak 23 22 Qpeak 17 09 RMS 16 23 AVG 11 36 C...

Страница 56: ...the Attenuator is now switched to 20dB automatically by 7010 MHT This query command MHT sends back a string expressing the HOLD TIME for manual mode Unit is fixed ms Example of reply MHT 1000 ms which...

Страница 57: ...is 10dB TMP This query command TMP sends back a string showing the temperature expressed in Celsius C Example of reply TMP 40 50 UHT This query command UHT sends back a string expressing the actual ho...

Страница 58: ...as been ignored for the following The instrument is not idling and therefore cannot switch to a different mode In such an event stop the current process Manual Analyzer or Sweep and put it back to Idl...

Страница 59: ...o Analyzer Reply section for details of the response The 7010 goes to Analyzer mode The 7010 goes to Analyzer mode Example SAGO SAHT h This setting command sets hold time for Analyzer mode The string...

Страница 60: ...in dB The procedure to creating a conversion factor on PMM7010 is the following Send as many commands as the frequency points of limit are in upwards order Terminate and make it active by sending the...

Страница 61: ...end the first line 150kHz 66 dB V SLDW 1 500e3 56 46 Send the second line 500kHz 56 dB V SLDW 2 5e6 56 46 Send the third line 5MHz 56 dB V SLDW 3 5e6 60 50 Send the fourth line 5MHz 60 dB V SLDW 4 30e...

Страница 62: ...end the fifth line 30MHz 60 dB V SLIE Custom CISPR 14 1 Send the name and make the limit active Note that all the higher indexes in the limit are cleared Therefore a limit must be made in upwards orde...

Страница 63: ...es to manual mode automatically Example SMAT 15 set attenuator to 15dB and disable Automatic Attenuator feature Example SMAT 1 Enable Automatic Attenuator feature SMHT h This setting command sets hold...

Страница 64: ...No more than 2 detectors can be indicated at once S Smart Mode this must be the first in the string Q Peak N C Rms A Avg R Rms C C Avg HoldTime Hold Time for each step and it is expressed in ms Note t...

Страница 65: ...SFD ERR 1 RC LF An Error was found in start or stop frequencies Errors can be one of the following o Start Frequency lower than minimum allowed frequency o Stop Frequency higher than maximun allowed...

Страница 66: ...minated by the command ASBK the end of sweep which will not contain all expected data is SBK OK RC LF Here is an example of the command SSFD 1e6 10e6 10e3 PAR 0 6 10 OFF ON Which triggers a sweep from...

Страница 67: ...tuned and measured as all remaining will be completely neglected Hence it is responsibility of the user to make sure the data sent are correct and coherent The reply is SSFW OK which acknowledges the...

Страница 68: ...he hold time to 2 ms SADT 2 Sets the detector to AVG More Command More Command SAGO Start Sweep in Analyzer Mode The PMM7010 replies by sending a 48 bytes header followed by a variable size array of i...

Страница 69: ...Fstart 2 Fstep Little Endian 16 bit integer representing the level referred to start frequency plus 2 step frequency 52 nn Level Fstop Little Endian 16 bit integer representing the level referred to...

Страница 70: ...the size of the array then a 512 bytes main header plus an additional 512 bytes sweeps header followed by a variable size array of integer the size depends on frequency settings as follows ID Byte Mne...

Страница 71: ...2 525 n 1 32 OffRec1 Little Endian 32 bit long representing the address of the array level of sub sweep 1 This figure must be multiplied by 2 526 n 1 32 527 n 1 32 528 n 1 32 529 n 1 32 LenRec1 Little...

Страница 72: ...step frequency of sub sweep n OffRecn 6 OffRecn 7 Level Fstart Fstep Alternate Detector Little Endian 16 bit integer representing the level of the alternate detector referred to start frequency of sub...

Страница 73: ...ain and sweeps header 4 Read the relevant data concerning Start Stop of full sweep as well as single sub sweeps if they are more than one 5 Go to the offset OffRec1 OffRec1 2 in the received buffer an...

Страница 74: ...with its FW It is used to check congruencies frequency limits and all the related characteristics S N Queries the serial number S N 000WE20304 Serial number CRA Queries if RMS AVG optional detector i...

Страница 75: ...n you would like to submit to the attention of our service organization is kept in great consideration Moreover we are continuously improving our quality but we know this is a never ending process We...

Страница 76: ...Suggerimenti Commenti Note Suggestions Comments Note...

Отзывы: