7-24
Remote control
7.7 Sweep Mode
commands
sequence
example
This is the sequence used by the PMM Emission Suite when a scantable is run, in
the “conducted” range (Band B from 0.15 to 30 MHz).
Example
Command sent Reply received
Description
#?IDN*
Queries information about model, release and date of the firmware
IDN=7010/03-FW - 1.09 11/06/14
It is a 7010/03 with its FW
It is used to check congruencies, frequency limits, and all the related
characteristics
#?S/N*
Queries the serial number
S/N=000WE20304
Serial number
#?CRA*
Queries if RMS-AVG optional detector is active
CRA=OK
RMS-AVG active
#SCFA -1*
Correction factors, if present, deactivated
CFA=OK
(OFF)
Important. Could be previously stored and activated.
#S3PRC*
Sets the unit in the bands A and B for conducted
3PR=OK
ack
#SSFD
150000;30000000;5000;P;2;6;10;OFF;OFF;0*
Sweep
Binary data reply
Measurements data
#?UHT*
Queries the hold time
UHT=2.5ms
Applied hold time