7-4
Remote control
7.6 PMM 7010
COMMANDs
Following are Commands to control and set various operating modes of PMM
7010, that can be distinguished in two main groups:
•
Query
Commands
•
Setting
Commands
7.6.1
Query
COMMANDs
Description
?3PR
This query command
#?3PR *
sends back a string expressing whether the
apparatus is a PMM 7010 and in which mode. The reply is either:
•
3PR =CON
(The PMM 7010 is ready for conducted frequency range )
•
3PR =RAD
(The PMM 7010 is ready for radiated frequency range )
•
3PR=NUL
(The PMM 7010 is not ready)
?AAT
This query command
#?AAT *
sends back a string expressing the Attenuator
status for Analyzer mode. The reply is made of 2 fields:
•
AAT =AUTO (
7010 automatically selects the most suitable attenuator to
get best dynamic range) or
AAT =MAN
( operator selects manually the
attenuator)
•
Value
(in dB) of current attenuator
Example of reply
: AAT =AUTO; 20
which means that the Attenuator is now
switched to 20dB automatically by 7010
?ACE
This query command
#?ACE*
sends back a string expressing the Center
frequency, in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply
: ACE = 1.07
which means that the Center frequency in
Analyzer mode is 15MHz
?ADT
This query command
#?ADT *
sends back a string expressing the detector used
for Analyzer mode. The reply cab be:
•
ADT =Peak
•
ADT =Avg
•
ADT =Rms
?AHT
This query command
#?AHT*
sends back a string expressing the HOLD TIME for
Analyzer mode. Unit is fixed ms.
Example of reply
: AHT= 1000 ms
which means that the hold time in Analyzer
mode is 1000ms (or 1s)
?AOP
This query command
#?AOP*
sends back a string expressing the Stop frequency,
in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply
: AOP = 3.07
which means that the Stop frequency in
Analyzer mode is 30MHz
?ART
This query command
#?ART*
sends back a string expressing the Start frequency,
in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply
: ART = 1.07
which means that the start frequency in
Analyzer mode is 15MHz
?ASP
This query command
#?ASP*
sends back a string expressing the Spanfrequency,
in exponential notation, for analyzer mode. Unit is fixed MHz.
Example of reply
: ASP = 3.07
which means that the Span frequency in
Analyzer mode is 30MHz
Содержание PMM 7010 Series
Страница 8: ...VIII Contents This page has been left blank intentionally...
Страница 36: ...4 6 Applications This page has been left blank intentionally...
Страница 50: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Страница 76: ...Suggerimenti Commenti Note Suggestions Comments Note...