Applications 4-1
4 – Applications
4.1 Measuring the
EMI Voltage
ElectroMagnetic Interference (EMI) voltage measurements on power
supply lines or on signal lines are carried out by means of "Coupling
Networks" (e.g. LISNs) or other transducers (e.g. antennas, voltage
probes). The frequency range is dictated by the applicable standard,
however it is generally limited from 9 kHz to 30 MHz in commercial
applications, while for measurements on other equipment/accessories -
e.g. military, automotive, information technology and communication
equipment (ITE), ISDN devices, etc. - the frequency range extends down
to 20 Hz and up to 200 MHz, depending the relevant standards.
Nevertheless, the most common tests in the frequency range covered by
PMM 7010 are certainly the conducted measurements, done with the help
of a LISN, and the ratiated ones, for which various types of antennas can
be used.
4.1.1 Measuring Principle
with a LISN
In the case of a system with two floating conductors, the EMI voltages of
the two conductors relative to each other and with reference to the ground
form a vector system where three kind of RFI voltages are present
(sometime EMI is also referred as RFI: Radio Frequency Interference).
They are:
•
Symmetrical (or differential mode) voltages. These RFI components are
measured between the two conductors. They behave like the wanted
signal on the forward and return lines.
•
Asymmetrical (or common mode) voltages. These EMI components are
measured between the electrical midpoint of the two conductor voltages
and reference ground. Usually these are the components most likely to
cause interference effects.
•
Unsymmetrical voltages. These RFI components are measured from
each line conductor and the reference ground. They consist of
symmetrical and asymmetrical components. The measurement of these
RFI voltages is the easiest to do and the most commonly performed.
Indeed, for practical reasons, Standards mostly specify the measurements
of unsymmetrical voltages instead of the theoretically more meaningful
common mode EMI, but the standard doesn’t necessarily reflect the real
life exactly as it is: it shall be a common practice reference aiming to create
a database of fully comparable results.
See figure 7-1 for a graphic explanations of these differences between the
EMI voltages.
Some Regulations require both measurement of symmetrical and
asymmetrical RFI voltages, also defining separate and different limits for
them.
4.1.2 Coupling Networks
Coupling Networks are electrical interfaces which allow to “transfer” the
quantity to be measured (e.g. EMI voltages or currents) from the lines
under test to the receiver.
Some types of Coupling Networks are: AMN (Artificial Mains Network),
also known as "LISN" (Line Impedance Stabilization Network), Current
Probe and Voltage Probe. Again, Fig. 7-1 also shows the AMN principle.
The PMM 7010 is already equipped with a built-in LISN.
Document 7010EN-81007-1.09 - © NARDA 2018
Содержание PMM 7010 Series
Страница 8: ...VIII Contents This page has been left blank intentionally...
Страница 36: ...4 6 Applications This page has been left blank intentionally...
Страница 50: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Страница 76: ...Suggerimenti Commenti Note Suggestions Comments Note...