7-12
Remote control
SLIM n
This setting command sets the Margin to be used with Smart Detector function. The
argument (n) in the range of -20 to 20.
The reply is
LIM =OK
which acknowledges the command has been granted or
LIM
=SERR
if the command has been ignored.
Example:
#SLIM 2*
Sets the threshold of smart detector to 2 dB below the limit.
SLIW n,freq;lev
This setting command is used for setting the custom limit frequency by frequency. It
consists of three fields as follows:
•
n
Is the index of frequency being written and it should range from 0 to 15
•
freq
Is the frequency, expressed in Hz (exponential notation allowed), being
written
•
lev
Is the level of the limit and is expressed in dBµV
The procedure to set a custom limit on PMM7010 is the following:
•
Send as many commands as the frequency points of limit are in
upwards order.
•
Terminate (and make it active) by sending the command
SLIE
Here is an example how to make a custom limit on PMM7010:
•
#SLIW 0, 150e3; 66 * Send the first line 150kHz 66 dBµV
•
#SLIW 1, 500e3; 56 * Send the second line 500kHz 56 dBµV
•
#SLIW 2, 5e6; 56 * Send the third line 5MHz 56 dBµV
•
#SLIW 3, 5e6; 60 * Send the fourth line 5MHz 60 dBµV
•
#SLIW 4, 30e6; 60 * Send the fifth line 30MHz 60 dBµV
•
#SLIE Custom CISPR 14-1 * Send the name and make the limit
active
Note that all the higher indexes in the limit are cleared. Therefore, a limit must be
made in upwards order otherwise all points, having a higher index previously written,
are cleared.
It is
responsibility of the user to make sure
the data sent are correct and
coherent.
The reply is
SLIW =OK
which acknowledges the command has been granted or
LIW
=SERR
if the command has been ignored.
SLSN n
This setting command
#SLSN n*
sets the input for Band B according to the
argument ‘n’ as follows:
•
0
Input on N connector
•
1
Input on 7010 internal LISN
L1
•
2
Input on 7010 internal LISN
L2
‘n’ higher than ‘2’ is ignored and SERR message sent back
The reply is
LSN =OK
which acknowledges the command has been granted or
LSN
=SERR
if the command has been ignored.
Example:
#SLSN 0*
switches the input for radiated to ‘N’ connector
Example:
#SLSN 2*
switches the input for radiated to the 7010 built-in LISN Line-2
SMAF f
This setting command sets the tuning frequency for manual mode. The string (f) can
be in exponential form and should be expressed in Hz. The reply is
MAF=OK
which
acknowledges the command has been granted or
MAF=SERR
if the command has
been ignored.
When the command is granted the 7010 goes to manual mode automatically
Example:
#SMAF150e3*
Sets the manual frequency to 150kHz
Содержание PMM 7010 Series
Страница 8: ...VIII Contents This page has been left blank intentionally...
Страница 36: ...4 6 Applications This page has been left blank intentionally...
Страница 50: ...6 8 PMM 9010 RMA Rack Mount Adapter This page has been left blank intentionally...
Страница 76: ...Suggerimenti Commenti Note Suggestions Comments Note...