© Ernst Nagel GmbH 2S Trimmer Service Manual en.doc Rev.1.0. 20.07.2011
116
Contents
6. Default / basic adjustments
(continued)
6.8 Tapping Transformer and blown fuse conditions
Fig.
6.61
Fig.
6.62
Fig.
6.63
Fig.
6.64
Removal/Adjusting
No. Action
Remarks
1
Remove Rear Cover
See Chap.
4.2
2
NOTE!:
Make sure the power cord of the STR is disconnected
from the wall outlet
3
Measure the wall outlet voltage (120V or 230V)
4
Check to which terminal the power cable is connected.
(The picture shows a transformer tapped for 230V)
See A
5
Tap the Transformer to 120V or 230V settings according to the
label on the Transformer by moving cable [A].
NOTE!:
Do not move cable [B].
(See Fig. 6.61 & 6.62)
6
Remove the current label ‘Disconnects only…’
(See Fig. 6.63)
7
Paste the correct label ‘Disconnects only BLTXX02 230VAC
50Hz’ or ‘Disconnects only BLTXX02 120VAC 60Hz’ depending
on the wall outlet voltage.
Included in Box with
accessories
(See Fig. 6.64)
Installation
No. Action
Remarks
1
For placing back Rear Cover follow the Rear Cover procedure in
reversed order.
See Chap.
4.2
Purpose of tapping Transformer
To be able to change the Transformer setting, depending on different voltages.
B
A