Troubleshooting for Problems Other than Error Messages.
If the procedures described below do not solve the problem, please contact your local dealer or J. MORITA
OFFICE.
Problem
Cause
Response
Does not turn on.
Circuit Protector may be
opened
Check the Circuit breaker, on the back of the unit. It may
be off.
Low laser output
or
Aiming beam is not
emitted
or
Aiming beam is
dim or hazy
Water leaks inside the
handpiece.
Water may seep inside the handpiece if the O-rings on the tip
or the handpiece are worn out.
This can interfere with the output of the laser or its aiming
beam or both.
Try using a different handpiece or a new tip.
Apply a little grease to the O-rings occasionally to keep
them from wearing out.
Tip may be worn out.
Wear or damage (chipping etc.) will lower the transparency
of the tip.
If the tip is worn down completely inside the metal sleeve, it
will not transmit the laser effectively and the laser output
will drop.
Replace the tip.
Poor assembly of hollow
waveguide
Take off the handpiece grips. Check the drum lens.
Make sure the hollow waveguide is lined up with the hollow
waveguide support. Loosen the ring for the hollow
waveguide and rotate the hollow waveguide until the Aiming
beam shines properly. Then retighten the ring.
Lens of Handpiece or R
handpiece grips is dirty.
Clean the drum lens for the handpiece and the ball lens for R
handpiece grips with a cotton swab occasionally. A dirty
lens will interfere with the laser output power and spoil its
focus.
Replace the lens if it is extremely dirty, scratched or
otherwise damaged.
Hollow waveguide is
broken
Contact your local dealer or J. MORITA OFFICE.
Spray is not emitted
from tip.
Spray water bottle is
empty.
Replace the spray water bottle.
No spray after replacing
the spray water bottle.
After the spray water bottle is replaced or after the unit has
not been used for a long time, there may be some air in the
water line. Put the unit in Ready mode and depress the foot
switch to its first level to run the water pump until spray
starts coming out.
Tip is plugged up
The tip may be worn out or plugged up. Or the water line
in the tip may be plugged. Replace the tip and see if spray
comes out normally.
Water connection for the
hollow waveguide may not
be properly inserted.
Reconnect the hollow waveguide. Plug it in until there is
an audible click.