Instructions for Use 2012-08-23
35
Refresh Flash Lamp
After considerable use, the flash lamp may no longer work so
well and cause errors to occur.
This procedure may rectify the problem. It takes about 15
minutes.
If the unit is use at low power for a long time, the terminals may
get dirty and interfere with ignition.
Operating the lamp at high power can clean the terminals.
The procedure stops after 15 minutes or you can stop it anytime
by pressing Cancel.
Check Software Version
Check the software versions for the control system, display, and
screen.