8. Troubleshooting
Explanation of Error and Caution Messages
Error and caution messages appear in the touch panel display. Follow the instructions which appear.
Contact your local dealer or J. MORITA OFFICE in the following cases:
•
Repairs are required
•
Replacement of parts such as the flash lamp, cooling water, deionization filter cartridge etc.
•
Updating the V-J table (calibrating output power)
•
Cleaning the internal filter
•
Frequent or repeated errors
A message appears in the operation panel when the following errors occur.
No.
Type
Explanation and Response
Reference
Interlock 1. Flash lamp defect.
Lamp is defective or doesn’t light up.
Response: Flash lamp is old and not working
properly. Go to Menu and use Refresh Lamp.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
Interlock 2 Main power supply is
abnormal.
Cannot charge Up.
Response: Restart the unit.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
Interlock 4 Cooling water problem
Cooling water is not circulating.
Response: Either the pump is not working or
there is no cooling water in the unit. Turn the
power off, wait about 10 seconds, and then turn
the power on again. Open the front cover and
see if there is enough cooling water.
Interlock 5 Shutter error
Shutter is not working properly.
Response: Restart the unit.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
Interlock A Hollow waveguide is
not connected.
The hollow waveguide is not connected.
Response: The hollow waveguide may be loose.
Tighten the connection ring and restart the unit.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
Interlock B Not enough cooling
water.
Not enough cooling water.
Contact your local dealer or J. MORITA OFFICE.
Interlock C Cooling water is too
hot
Cooling water is too hot, over +45°C (+113°F).
Response: Wait until the water cools down to
below +45°C (+113°F). Check current
temperature.
This will happen less often if there is plenty of
open space in back of the unit.
If this happens
frequently, the filter
inside the unit may be
plugged up. Contact