Instructions for Use 2018-04-23
43
2) Autoclaving (sterilization) (Always perform this procedure after cleaning and before use)
The autoclaving process is intended to destroy infectious microorganisms and pathogens.
(1) Put the handpiece grips, hanger and tips in sterilization
pouches or the tip stand for autoclaving.
♦
The contact tip could be damaged if it is sticking up
when the top of the stand is closed.
(2)
Place them inside the autoclave chamber.
♦
Tips are easily broken. Take care that tips do not
bump against each other or agianst other instruments
when putting them into the autoclave. Do not drop or
bump them against anything when handling them.
(3) Set autoclave cycle to the following parameters:
Temperature: +135°C (+275°F)
Time: 10 to 15 minutes
Dry Time: 0 min (dry naturally)
♦
Do not use any washer disinfectors as well as Class B
Autoclave equipments.
♦
Do not use the drying stage if the autoclave has one.
The temperature might be too high.
(4) At the time of completion of the autoclave, let them cool.
♦
Make sure all 3 O-rings of tips are intact and not
damaged in any way. Pay special attention to the
smallest one. If this is missing or damaged, water
could seep into the handpiece and damage it or cause
the laser to lose power. It also might harm the drum
lens.
Three O-rings
Medium
Large
Small