Instructions for Use 2018-04-23
19
(3) Handpiece and Contact Tip Attachment
(1)
Hold the ring in one hand and then put the handpiece Grips
on by turning it until it clicks into place.
* To remove it, hold the ring and pull it off.
♦
If you do not hold the tip end of the handpiece Grips and
turn when you put it on, the O-ring will be damaged.
♦
Grease the end of the handpiece periodically to prevent
damage to the O-ring.
(See page 45.)
(2)
Hold onto the outer cover of the hollow waveguide and
give the handpiece a light tug to make sure it will not
come off.
(3)
Make sure the contact tip is clean and free of blood and
other contaminants. Make sure all 3 O-rings are in place.
(4)
Grip the knurled part of the contact tip and screw it onto the
handpiece Grips.
♦
Only hold the knurled grip. Never grip the pipe; this
could damage the tip.
♦
R200T and R300T tips are required to attach to the R
Handpiece Grips.
Outer Cover
Handpiece Grips
Ring
Knurled Grip
Three O-rings
Medium
Large
Small