Instructions for Use 2018-04-23
21
(4) Operation Procedure
1) Set Laser Irradiation Conditions
(1)
Energy Level
•
Press the “mJ” part of the panel; a window to make
this setting will appear.
•
Press a number to make the setting.
•
You can also press the plus and minus signs.
For less than 100 mJ, values can be set in 5 mJ steps.
For more than 100 mJ, values can be set in 10 mJ steps.
Setting Ranges: 10pps — 30 mJ to 400 mJ
20pps — 30 mJ to 170 mJ
25pps — 30 mJ to 80 mJ
Press “Close” after making the setting.
PPS means pulses (laser shots) per second.
•
Press one of the numbers to make this setting.
•
The mJ display turns red if mJ is set at 150 or higher.
(2)
PPS (pulses per second) Setting
•
Press the “PPS” part of the panel; a window to make
this setting will appear.
•
Press one of the numbers to make this setting.
(3 PPS is actually 3.3 pulses per second.)
The total amount of energy delivered in 1 second can be
found by multiplying the energy level setting by the setting
for the pulse rate.
(3)
Turn Aiming Beam On or Off
A red aiming beam is emitted from the handpiece.
•
Press the Aiming key to turn the aiming beam on or off.
•
The aiming beam is turned on when the unit starts up.
* The Aiming Key will be lit up.
•
Turn the aiming beam off if you don’t need it.
•
The aiming beam is emitted when the unit is in Ready
mode as well as during laser emission.