Instructions for Use 2018-10-22
44
(2) Cleaning the Main Unit
Prevent contagion and contamination by cleaning the unit regularly.
Wipe the outside of the main unit with ethanol for disinfection (Ethanol 70 vol% to 80 vol%) or a neutral
detergent.
Do not use ozone or ultra violet light to disinfect the treatment area. This could damage plastic, rubber or other
materials.
•
Use only ethanol for disinfection (Ethanol 70 vol% to 80 vol%) or a neutral detergent. Alkaline and acidic
cleaners, liquid cresol soap,and other chemicals may damage or discolor the surface. Do not use solutions that
contain cresols, triclosans,hypochlorite,aldehydes,or quaternized ammonium salts. (Check the ingredients for
disinfectants.)
•
Immediately wipe off any chemicals that are spilled on the unit with ethanol for disinfection (Ethanol 70 vol%
to 80 vol%).
•
Do not press down too hard to wipe the surface; this could cause peeling.
•
Prevent contagion and contamination by cleaning the unit regularly.