Interlock D Cooling water too cold
Cooling water is too cold, less than +15°C (+59°F).
Wait for it to warm up.
Response: Leave the unit on and wait for the
water to warm up; it will then automatically start
to operate normally. Check current temperature.
This commonly
happens in the winter
when the room is cold.
Interlock F Cover Interlock
Cover interlock activated.
Response: Restart the unit.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
Interlock G Remote Interlock
Remote interlock activated.
Response: Check the door for the remote
interlock. Or check the remote interlock
connection on the back of the unit.
100
Emergency Stop Alarm
The emergency stop switch has been pressed.
Response: Turn off the main power and release
the emergency switch. Then restart the unit.
Push the emergency
switch again after it has
been activated to
release it.
101
Watch Dog Timer
Watch Dog Timer activated.
Response: Restart the unit. If this does not
work, contact your local dealer or J. MORITA
OFFICE.
102
Switch error
A switch error was detected when the unit was
turned on.
Response: This happens if the foot switch is
depressed when the unit is turned on. Let the
foot switch up and restart the unit.
The foot switch is
checked for safety
when the unit is turned
on.
103
Memory Back-up
Error (SRAM)
Battery for memory is low. To recharge the
back-up battery, turn on the power and leave it on
for 30 minutes. Then reset the clock and rewrite
the names for the memories.
A rechargeable battery
is used to maintain the
clock and other
functions. Turn the
unit on once every 6
months to recharge the
battery.
104
Laser output Error
Laser output does not match set value.
Response: The flash lamp is probably old and not
working right. Go to Menu and use Refresh
Lamp.
If this does not work, contact your local dealer or J.
MORITA OFFICE.
105
Energy Setting Error
Energy level cannot be properly set.
Response: Probably needs calibration. Contact
your local dealer or J. MORITA OFFICE.
This happens if the laser
has not been calibrated
for some time.
106
Voltage limit stop
Cannot produce the output power that has been
set.
Response: Lower the output power (mJ), or
replace the flash lamp.
If an error occurs even after the power has been
lowered, a mirror may be damaged; in this case,
contact J. MORITA OFFICE.
This happens if the
flash lamp is in poor
condition
110
Temporary power
failure
Temporary power failure error.
Response: Restart the unit. Check the socket for
the main power cord.
Happens when main
AC power source is
temporarily lost.