Instructions for Use 2018-10-22
23
•
Take great care when using the tip air inside a body cavity or tubular lumen. Raising the air pressure
inside a cavity or lumen could force air into a blood vessel through an open wound and result in a gas
embolism. Also take great care when using the tip air in areas of the oral cavity where it could
increase the pressure; this could result in a severe gas embolism or subcutaneous emphysema.
•
Never look directly in the aperture where the guide light comes out or let it strike anyone in the eye;
this could result in blindness.
•
Irradiating hard tooth tissue without using spray could cause carbonization. When irradiating hard
tissue, make sure the spray is turned on and that enough water is being delivered to the treatment area.
•
Do not set output powers greater than that specified for the contact tip; this could overheat the tip.
•
Before using the laser, check if the spray is on or off and what the volume adjustment is. Depress the
foot switch to the first level only to check the spray.
•
Make sure tip air flow is not so great that it damages tissue.
•
If an error is indicated in the operation panel, stop operating the laser unit at once.
•
When the spray water setting is off, the contact tip temperature may rise up to +50°C (+122°F). Do not
allow the tip to make contact with body tissue for more than 1 minute.