Functional Safety Dual Input Smart HART® Temperature Transmitter
STZ
SECTION 5
Figure 5.34.
STZ Custom Calibration Menu - AO Loop Current Test/Trim
Two Point
1. Navigate to the “AO Loop Current Test/Trim” page.
2. Select the Trim Mode (two point).
3. Click on the Apply button.
4. Send the Store to device command.*
5. Click on the appropriate button for the AO level being trimmed (Lower or Upper Output
Range).
Note
: This puts the AO in manual mode and the actual AO value is given.
6. Enter the given Analog Output value in mA in the “Measured Output Current” entry field.
7. Click on the “Trim” button.
8. Repeat steps 6-7 until the desired value is obtained.
9. Repeat steps 5-8 for the second point to be trimmed.
10. If you are done trimming, click on the “Clear” button.
Note
: If you do not click the Clear button, the Analog Output will remain in manual mode and
will NOT follow the input signal.
11. You are done with two point trimming.