20
Molift Air / www.molift.com
Emergency Stop/Lowering
Remove Emergency stop relay and button
1. Activate emergency stop
2. Remove both side covers and disconnect
battery
3. Remove battery clamp (hex key 5mm) and
battery.
4. To access the relay with standard tool it is nec-
essary to remove screws holding gear/motor to
lower cover, and remove microswitch.
5. Remove nut and washers from emergency stop
relay with 14mm wrench.
6. Disconnect wires and remove the relay and the
button.
Emergency Stop/Lowering Parts
Article no.
1101588 4 mm red Polyester/cotton cord
1101587 Knob red (Cord End)
2510003 Emergency Stop Knob
2510002 Emergency Lowering Switch
2510099 Tube 6/8 mm red Length 40 cm
Check after installation
Activate emergency stop by pulling cord.
Make sure it is not possible to run lifter
with emergency stop activated.
Pull and hold cord. Electric emergency
lowering should start.
Perform and fill out Safety checklist
after service and repair