18
Molift Air / www.molift.com
Angle Sensor / End sensor
The hoist has a directional safety feature that
prevents lifting if angle on lifting band is to steep.
>30˚
If the hoist doesn’t respond to Hand control “up”
button, it might be because lifting band angle is
to steep, or lifting band is twisted and sensor is
activated.
This sensor also detects the end of the band, and
stops the lifter when the band is driven all the way
in.
Change angle sensor plastic part
1. Open sidecover and disconnect battery
2. Remove bottom (see “Bottom Plate” page 16)
3. Remove suspension coupling (see “Suspension
coupling page 19)
4. Change angle sensor plastic part, make sure to
install it the correct way (chamfered edges in
center facing up).
Change angle sensor microswitch
1. Open sidecover and disconnect battery
2. Remove bottom cover (see “Remove bottom
plate” page 16)
3. Use a plyer to pull the broken microswitch out,
or push it out from the inside of the lifter.
4. Cut the wires, and pull the wires out through
the hole.
5. Solder the wires to the microswitch on the
outside and use heat shrink tube to isolatee.
6. Push the microswitch into the hole to lock it in
place.
7. Mount bottom plate.
Check after installation
Run the lifter, and pull band sideways to
activate angle sensor. The lifter should stop
lifting when angle sensor activates.
Drive the band completely in. The sensor
should aktivate and lifter stop running
when the band is driven all the way in.
Perform and fill out Safety checklist
after service and repair
Angle sensor Parts
Article no.
2510006
Angle sensor plastic part
2510007 Microbryter Series 1050 Wlfa
35-830-55
Содержание Air 300
Страница 1: ...TM16101 Rev A 2013 11 14 Molift Air EN Technical manual...
Страница 8: ...7 Molift Air www molift com Air Basic Exploded Drawing 20 001 Rev K Air Components...
Страница 10: ...9 Molift Air www molift com Air IRC Basic Exploded Drawing 20 012 Rev E...
Страница 35: ...34 Molift Air www molift com Notes...