2. Open the injector pump access door.
Item
Name
1
Retaining stop
2
Pump wheel
3
Injector pump access door
Injector Pump With Access Door Opened
3. To remove the tubing from the pump housing, pull the retaining stop away from the
injector, and then disconnect the ends of the tubing from the barbed fittings.
4. Loop the portion of the tube located between the retaining stops around the pump
wheel. Stretch the pump tubing so that it is held in place by the retaining stops. Make
sure the tube is completely around the pump wheel.