Chapter 2: Setting Up the Instrument
0112-0174 C
17
8. If the instrument configuration includes an injector module, insert one end of the
injector module power supply into the injector module power port on the back of the
injector module and the other end into the power cord. Then plug the power cord
into a grounded electrical outlet.
9. Insert the square end (release tab up) of the injector module communication cable
into the PORT on the back panel of the instrument and then insert the round end
into the COMM IN port on the rear of the injector module.
Item
Name
1
PORT - Injector module communication port on instrument
2
COMM IN - Communication in port on injector module
Communication Cable Connection for Injector Module
10. Turn the instrument around so that the control panel faces you. Ensure no cables run
beneath the instrument. Leave sufficient space between the back of the instrument
and the nearest objects or surfaces to enable the ability to lift the instrument hood
and injector hood and to ensure proper ventilation and cooling.
Assembling the Injectors
The base instrument configuration does not include injectors. The instrument
configuration can have several factory installed injector setups. For all instrument
configurations that include injectors, the wash bottles and reagent bottles remain
outside the instrument.
When the instrument configuration includes two injectors there is one pre-injector (P-
injector) and one measurement injector (M-injector). The injectors are built into the
instrument with injector ports located on the left side of the instrument.