APPENDIX E – ALARMS
83
CODE
POSSIBLE CAUSE
PROPOSED SOLUTION
Steam leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again.
Excessive load.
Perform RESET following the instructions.
Check the quantity of material in the sterilization chamber and make sure
it does not exceed the maximum quantity allowed.
(See the Summary Table in Appendix A, Technical Characteristics).
Problem in the plumbing circuit.
Contact the Technical Support Department (see Appendix Z).
Intervention of the steam generator
safety thermostat.
Heating element safety thermostat
intervened.
A 252
Heating or steam generator heating
element malfunction.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z).
Presence of water or condensate in
the sterilization chamber.
Perform RESET following the instructions.
Carefully dry the inside of the sterilization chamber and start the program
again.
Do not put material impregnated with water, or liquids in general, in the
chamber.
Air leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again.
Vacuum pump broken.
A 253
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z).
Water injection pump malfunction.
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z).
Intervention of the steam generator
safety thermostat.
Heating element safety thermostat
intervened.
A 254
Heating or steam generator heating
element malfunction.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z).
Steam leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again.
Excessive load.
Perform RESET following the instructions.
Check the quantity of material in the sterilization chamber and make sure
it does not exceed the maximum quantity allowed.
(See the Summary Table in Appendix A, Technical Characteristics).
Problem in the plumbing circuit.
Contact the Technical Support Department
(see Appendix Z).
Intervention of the steam generator
safety thermostat.
Heating element safety thermostat
intervened.
A 255
Heating or steam generator heating
element malfunction.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z).