5. FIRST START-UP
18
MAX LEVEL IN THE
INTERNAL/
EXTERNAL DRAIN
TANK
Emptying the internal
tank
Emptying the external
tank (option)
When the water level in the internal or external drain tank reaches the MAX level, the LCD
display alternatively lights the MAX and MIN icons.
NOTE
I
N THIS CONDITION THE UNIT WILL GENERATE AN ALARM INDICATION
(
SEE
A
PPENDIX
E
-
A
LARM
)
AS YOU ATTEMPT TO LAUNCH A STERILIZATION CYCLE
.
In this case, empty the internal or external draining tank.
Referring to the figure, open the door and operate in the
following way:
1. Arrange an empty tank on the floor near the sterilizer
and put the free end of the supplied tube into the tank;
2. Connect the quick connector to the corresponding
female connector under the chamber entrance
(marked
), pushing until you hear a click;
3. Wait for the complete empty of the internal tank; then
while pinching the tube with your fingers, press the
metal lever located on the side of the connector and
detach the quick connector.
Remove the top cap from the external tank and empty into a sink the water exceeding the
signed level.
WARNING
DO NOT EMPTY THE TANK COMPLETELY, BUT KEEP A QUANTITY OF
WATER UP TO THE MARKED LEVEL. OTHERWISE THE WATER DRAINING
SOUND AND THE STEAM ESCAPE FROM THE VENT-HOLE WILL
INCREASE CONSIDERABLY.
Refer to chapter “CONNECTING AN EXTERNAL DRAINING TANK” for more details.