APPENDIX E – ALARMS
76
Alarm outside
the cycle
At the end of what has been described and having reached safe conditions, the machine
activates a special procedure, that asks the user to manually unlock the door:
P R E S S
Ç
T O
U N L O C K T H E D O O R
1 0 1 . 0 ° C X X X X
0 . 0 1 b a r 1 1 : 4 0
Press the
Ç
key to unlock the door lock mechanism; the following message appears:
( A l a r m M e s s a g e )
O P E N T H E D O O R
1 0 0 . 8 ° C X X X X
0 . 0 1 b a r 1 1 : 4 2
Once the door is open, the user is finally asked to reset the system:
( A l a r m M e s s a g e )
R E S E T S Y S T E M
9 5 . 5 ° C X X X X
0 . 0 0 b a r 1 1 : 4 3
Perform a RESET (described below) and then turn-off the equipment and check the error or
make the repair.
NOTE
W
HEN THE DOOR IS OPENED
,
THE REPORT
(
NORMAL OR EXTENDED DEPENDING ON THE
TYPE OF ALARM
)
WILL BE PRINTED FOR THE INTERRUPTED STERILIZATION PROGRAM
AND THE ALARM THAT INTERVENED
.
C
HECK THE DOCUMENT
,
INITIAL IT IN THE SPACE
PROVIDED AND FILE IT IN A SUITABLE PLACE
.
R
EFER TO THE PRINT REPORT EXAMPLES
SHOWN IN
A
PPENDIX
B,
P
ROGRAMS
”.
If the alarm intervenes outside the sterilization or test program the display will show:
( A l a r m M e s s a g e )
A L A R M !
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Turn-off the equipment and check the alarm.
Or, depending on the type of alarm:
( A l a r m M e s s a g e )
W A R N I N G !
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Alarm Message
Alarm Code
Alarm Message
Alarm Code