APPENDIX B – PROGRAMS
58
NOTES
1) F
RACTIONATED
=
PRE
-
VACUUM WITH THREE VACUUM PULSES
(
SEE FIGURES IN THE FOLLOWING PAGES
)
S
INGLE
=
PRE
-
VACUUM WITH SINGLE VACUUM PULSE
(
SEE FIGURES IN THE FOLLOWING PAGES
)
2) L
ONG
=
10
MINUTES VACUUM DRYING
(
TYPICAL OF
POROUS
AND
WRAPPED
CYCLES
)
S
HORT
=
4
MINUTES VACUUM DRYING
(
TYPICAL OF
HOLLOW
AND
SOLID
CYCLES
)
3) A
CCESS TO A
CUSTOM
CYCLE DOES NOT REQUIRE A PASSWORD
.
N
ONE OF THE COMBINATIONS POSSIBLE IN
THE CUSTOMIZATION PHASE CREATE ANY RISKS OR DANGERS OF INJURY TO THE OPERATOR OR DAMAGE TOT
HE DEVICE
.