6. CONFIGURATION
29
Setting the STAND-BY
mode
(STAND-BY OPTIONS on
the ADVANCED menu)
Based on the equipment's frequency of use, or other considerations, it is possible to select the
heating level during the STAND-BY (preheating) phase and the time beyond which STAND-BY
is deactivated.
When you select STAND-BY OPTIONS with the
↵
key, you access the following menu:
→
S T - B Y M O D E +
S T - B Y T I M E - O U T
↑
E X I T
↓
-
When you select STAND-BY MODE, an additional menu appears where you can set the
heating level:
→
O F F +
L O W
↑
H I G H
↓
E X I T -
Select HIGH (high preheating level) for intense use or, at any rate, to reduce the wait time
between one cycle and the next to a minimum.
Select LOW (low preheating) for normal use, since the wait time will be relatively shorter, in
any case.
Select OFF (deactivate preheating) for occasional use. In this case, the wait time will be longer
(up to about 10-12 minutes for a "cold start").
Move using the + and - keys; confirm with the
↵
key.
On the other hand, when the ST-BY TIME-OUT option is selected, it is possible to set the time
for deactivating STAND-BY, i.e., how many minutes after the last cycle the heating elements
are turned off.
The following screen appears:
T I M E O U T : X X X
m i n
+ / - t o s e t
↵
t o e n t e r
Ç
t o e x i t
It is possible to set a value between 0 and 300 minutes (in 30-minute increments), after which
the heating elements are turned off (a condition analogous to STAND-BY OFF), avoiding the
useless consumption of electricity.
Set using the + and - keys; confirm with the
↵
key.
NOTE
T
HIS OPTION IS ALSO ACTIVE WITH
STAND-BY
OFF.
H
OWEVER
,
IN THIS CONDITION
THE TIMER VALUE OBVIOUSLY HAS NO EFFECT SINCE THE HEATING ELEMENTS ARE
TURNED OFF ANYWAY AT THE END OF THE STERILIZATION PROGRAM
.
W
HEN ANY CYCLE SELECTION KEY
(
STERILIZATION OR TEST
)
IS PRESSED
,
OR THE
MACHINE IS TURNED OFF AND ON WITH THE MAIN SWITCH
,
THE ORIGINAL
STAND-BY
MODE
(HIGH
OR
LOW)
IS IMMEDIATELY REACTIVATED
.