11. TEST PROGRAMS
51
Close the door and start the program with the START key.
NOTE
I
F A PASSWORD HAS BEEN SET WITH THE
ANY
CYCLE
START
OPTION
(
SEE THE
C
HAPTER
,
C
ONFIGURATION
,
S
ETTING THE PASSWORD
),
YOU WILL BE ASKED TO ENTER
THE ACCESS CODE
.
I
N ADDITION
,
THE EQUIPMENT CHECKS THE PRINTER PAPER PRESENCE AND
,
IF
MILLFLASH
IS CONNECTED
,
THE PRESENCE OF THE
C
OMPACT
F
LASH CARD AND ITS
MEMORY CAPACITY
.
T
HE POSSIBLE WARNING MESSAGES
,
AND THE CONSEQUENT ACTIONS TO CARRY OUT
,
ARE THE SAME AS DESCRIBED FOR A STANDARD STERILIZATION CYCLE
.
The vacuum phase begins immediately and the display reads:
V A C U U M T E S T
V A C U U M P U L S E
- 0 . 6 9 b a r 0 1 : 0 2
The display shows the pressure (bar), and the total time from the start of the program.
When the pre-set pressure is reached (-0.80 bar) the pump stops and the pressure
stabilization phase begins (WAITING PERIOD), which lasts 5 minutes (shown on the display
as a scalar value):
V A C U U M T E S T
W A I T I N G P E R I O D
0 5 : 0 0
- 0 . 8 0 b a r 0 1 : 4 8
During this phase, a variation of the maximum low pressure is allowed of not more than 10%,
without this causing the test to fail.
When the wait phase ends, the pressure verification phase, true and proper, begins
(LEAKAGE PERIOD), with a duration of 10 minutes:
V A C U U M T E S T
L E A K A G E P E R I O D
1 0 : 0 0
- 0 . 7 9 b a r 0 6 : 4 8
In this phase, a variation of up to ±0.02 bar is allowed, compared to the initial phase value.
Higher variations cause the test to fail.
The time is counted down until the phase is completed.
When this phase is also completed, the pressure is brought back to atmospheric pressure.
V A C U U M T E S T
L E V E L L I N G
- 0 . 2 9 b a r 1 7 : 1 9