9. RUNNING THE PROGRAM
45
Open the door
Report print
Equipment ready
NOTE
A
T THE END OF THE CYCLE
,
AND UP TO THE OPENING OF THE DOOR
,
THE HEATING
ELEMENTS ARE OFF
.
A
S A CONSEQUENCE
,
THE DEVICE IS SLOWLY COOLING
REGARDLESS OF WHAT THE
STAND-BY
MODE IS
.
NOTE
W
HENEVER THE STERILIZER
'
S
'
DOOR IS NOT OPENED AT THE END OF THE CYCLE
,
THE
VACUUM PUMP IS PERIODICALLY ACTIVATED TO REMOVE ANY TRACES OF CONDENSATE
FROM THE STERILIZATION CHAMBER
.
T
HE DISPLAY SHOWS
:
F O R C E D V E N T I L A T I O N
Ç
t o s t o p
3 5 . 2 ° C
- 0 . 0 2 b a r 2 9 : 4 0
Press
Ç
to interrupt ventilation and open the door.
Open the door and retrieve the sterilized material, using the extractor provided.
The icon
symbol goes off.
When the door is opened, the device goes to STAND-BY mode as previously set.
When the door is opened, the report for the sterilization cycle executed is automatically
produced. Check the document, initial it in the space provided and file it in a suitable place.
Refer to the print report examples shown in Appendix B, Programs.
NOTE
I
F SELECTED THE PRINTOUT
STEP
BY
STEP
OPTION
,
THE REPORT WILL BE PRINTED
DURING THE PHASES OF THE CYCLE
.
The device is ready to execute a new cycle.
Repeat the procedures explained in the Chapter, “Program Selection” for executing a new
sterilization cycle.
RESULT OF THE
CYCLE
After the cycle is finished, it is important to check the sterilization results.
Whenever a cycle finishes (message CYCLE COMPLETE and icon
on ), without,
therefore, being interrupted by any type of alarm, you are guaranteed to have completely
aseptic material.
The report of the sterilization parameters is an additional check tool.