Remove the fill tube cover at the bottom of the dewar.
Fill tube
O-ring
Fill tube cover
•
Carefully slide the fill tube into place through the opening of the top cover of the
MultiGas,
CAUTION:
Take care not to cross thread the Dewar adaptor plate threads and the
optics box threads.
•
Carefully screw the Dewar into the top of the optics box.
•
Rotate the Dewar slowly as to not strip the adaptor plate.
•
This process should only be hand tight, making sure you compress the
o-ring.