MKS MultiGas High-Speed Gas Sampling Cart
Connect
Sample
line
1.
Attach the ½” Swagelok quick connect (shipped with the cart) to the heated sample line.
Note:
It is suggested to use a 3/8” or ½” sample line with the High Speed cart
Note:
It is suggested not to exceed 35 ft. of sample line
Note:
It is suggested to use a heated stainless steel sample line
Note:
Be sure to minimize the amount of potential cold spots in the sample train.
2.
Locate the opening at the rear of the heated box.
Quick connect release
Heated line and quick
connect seated
View into the heated box
3.
Insert the heated line and Swagelok connecter into the opening until it clicks.
o
If you need to remove the heated sample line, push the quick connect release.
CAUTION:
If the sample line is not correctly seated no flow will go to the pumps.
When the pumps are turned ON they will dead head.
Installation and Use Manual 2011 APRIL 2 Rev. 1.5