Page 13
SPECIAL USER ATTENTION
NOTE
The table top must be centered or slid
toward the foot end for the Beach Chair
function to operate.
CAUTION
Please use the emergency switch at the
urgent case only.
CAUTION
The safety interlock system is not op-
erational when the emergency back-up
control switches are used.
CAUTION
The EMERGENCY BRAKE LOCK
switch does not activate the brake sys-
tem timer. The switch must be held until
the brakes are completely locked, ap-
proximately 10 seconds.
NOTE
The emergency back-up control
switches will function when the table is
operating on 100 - 240 VAC power,
battery power, or turned off.
NOTE
x
The Emergency Brake Release Valve
must be closed and tightened (coun-
ter-clockwise) before activating any
function.
x
If the Emergency Brake Release Valve
has been operated, the BRAKE
UNLOCK button on the pendant control
will have to be pressed before brakes
will lock again.
NOTE
To make the Back Section easier to
handle, remove the Head Section, pad
and X-ray top prior to removing the
Back Section. Remove the pad and
X-ray top prior to removing the Leg
Section.
WARNING
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and
cardiac defibrillator monitors. Improper
operation procedures might cause a
burn on an operators or patient, and
also cause the malfunction of the
product.
WARNING
x
When a patient is put on a mattress and
his/her position is set up, a patient shall
be set up to more than 1 cm apart from
a side rail so that a patient dose not
touch on side rails (metal).
x
The mattress should be attached in
order not to come off while being in use.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
WARNING
The antistatic properties of the table are
dependent on the use of the OEM pad
set which was offered with the table or
an alternate approved replacement.