SECTION II OPERATION
d. Back Section
To raise the back section, press
the BACK UP button (figure 2-9). The back section
will raise up to 90° above horizontal. To lower the
back section, press the BACK DOWN button. The
back section will go down to 40° below horizontal.
NOTE
If the table top is slid toward the foot
end, the back section will not go below
horizontal. An audible alarm will sound.
NOTE
To prevent damage to the kidney lift, a
safety interlock prevents the back sec-
tion from going more than 45° above
horizontal if the kidney lift is not all the
way down. An audible alarm will sound.
BACK UP
BACK DOWN
90°
40°
Figure 2-9. Back Section Positioning
e. Elevation
To raise table top, press the TABLE
UP button (figure 2-10). The table will lift a patient
weight of 450 kg up to a maximum height of 1100
mm (1205 mm with X-Ray top and 50 mm pad). To
lower the table top, press the TABLE DOWN button.
The table top will go down to a minimum height of
620 mm (minus pad).
CAUTION
The combination of minimum elevation
(top all the way down), extreme
Trendelenburg positioning and top slide
function may allow the back section to
collide with the base or floor.
TABLE UP
TABLE DOWN
1100 mm
620 mm
Figure 2-10. Elevation Function
f. Top Slide
To move the table top toward the
head end, press the SLIDE HEAD button. From
center position, the top will slide up to 245 mm. See
figure 2-11.
NOTE
If the leg section is positioned more
than 45° below horizontal, the top will
not slide toward the head end beyond
center. An audible alarm will sound.
NOTE
If the leg section is below horizontal,
slide toward head is limited to 195 mm
from center.
CAUTION
The combination of minimum elevation
(top all the way down), extreme
Trendelenburg positioning and top slide
function may allow the back section to
collide with the base or floor.