SECTION IV ELECTROMAGNETIC EMISSIONS
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
The use of Accessories, transducers, and cables other than those specified, with the exception of transducers
and cables sold by the Manufacturer of this device as replacement parts for internal components, may result in
increased Emissions or decreased Immunity of the 3602X Surgical Table.
The 3602X Surgical Table should not be used adjacent to or stacked with other equipment and that if adjacent
or stacked use is necessary, the 3602X Surgical Table should be observed to verify normal operation in the
configuration in which it will be used.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The 3602X Surgical Table is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the 3602X Surgical Table should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The 3602X Surgical Table uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The 3602X Surgical Table is suitable for use in all
establishments, other than domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies