Page 36
If the quality coefficient should be outside of the recommended valid
range, the measurement result is questionable:
Was the input of the reflector type correct? If not, the result
will be wrong.
Has reflector possibly been damaged or destroyed?
Foils may be damaged or destroyed during placement by the
construction machines. Depending on the extent of damage,
the reason for measurement failures could be foils with bent
ends, torn foils or perforated foils. Therefore, whenever
possible, always use robust reflectors for electromagnetic
thickness measurements!
Is the installed reflector certified for use with MIT-SCAN-T3? If
not, the signal may not match with MIT-SCAN-T3 calibration
files.
In this case, the result of the calculated layer thickness cannot
be guaranteed.
Evaluation of
measuring site via
material coefficient:
Reflector selection
correct?
Reflector of verified
quality used?
Reflector damaged?
Only robust reflectors of
verified quality ensure
reliable measurement
results.