35
5
−
2
−
4. Zero fine adjustment
(1) Procedures by key operation
●
During the Tare weight cancellation(A/Z) or Zero set is executed, and
also during effective for Zero tracking and Peak ON, Zero fine
adjustment mode can’t be entered into. (Displays ER
−
5.)
The Zero fine adjustment can be entered after Tare weight
cancellation clear(A/Z OFF), cancellation of the Compensated data at
Zero set (Execution of F
−
98), set OFF the Zero tacking (Set the
F
−
08 and F
−
09 to “00000”.) and set the Peak OFF.
●
During the calibration procedure, press the
key to interrupt the
calibration. The calibration data will keep the same condition as it is
entered before, then returns to the Measurement mode.
●
When the
key is pressed with the load display of “FUNC”, the
display will change as the following arrow marks indicate at every
time the key is pressed. However, every time the
is pressed, the
display will change as the reverse direction of the following arrow
marks.
“FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”
→
“SPAN”
→
“TARE”
→
“CHEK”
→
“MONT”
→
“VCAL”
→
“VADJ”
→
“FUNC”
→
“CCAL”
→・・・・・
(Hereinafter, it will repeat.)
Procedures
1
Press the
key for about one second.
The load display will show “FUNC”.
2
Press the
key four times.
It will make the load display proceeded as
“FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”.
Here, set the initial load application.