14
4
−
3
−
3. Connection with contact outputs
Connections with contact outputs “S1” and “S2” and external load should be made by using
each terminal and terminal No. 13 at “COM.2”.
At the same time, take care that the load should not exceed the rated load of contact output.
Rated load of contact output
AC125 V
0.1 A (Resistance load)
DC30 V
0.5 A (Resistance load)
〜
Shield
load
Surge preventive element
or
or
Warning
●
Connections with contact outputs should be made securely according
to the figures and also within the rated capacity of the instrument. If
neglected, it may cause an unexpected failure and/or malfunction to
the instrument.
●
For the protection from the contact of the instrument, connect the
surge preventive element that satisfies the characteristics of external
load to connect. If neglected, it may cause unexpected failure and/or
malfunction due to the effects from damage/melt down of the contact
and so on.
●
For the connections with contact outputs, be sure to apply shielded
cable, and the shielded cable should be connected with GND terminal
of the instrument.(Terminal No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.