Procedures
4
Press the
key.
The load display will flash on and off showing
the current load value.
At the same time, adjust the deviation with the
right keys so the analog output will meet with
the actual weight.
It can be increased continuously by the
continuity press of the key.
key
:
Decrease the analog output
key
:
Increase the analog output
5
Fine adjustment on the maximum value of
analog output
Press the
key.
The load display will show “A_HI”.
At the same time, set the weight equivalent to
the maximum output value of the analog output.
6
Press the
key.
The load display shows the current load value
and flashes on and off.
At the same time, adjust the deviation of analog
output against the load with the right keys.
It can be increased continuously by the
continuity press of the key.
key
:
Decrease the analog output
key
:
Increase the analog output
7
Press the
key.
The load display will show “END”.
By pressing the
key again, the
Measurement mode can be returned through the
VADJ mode, then the present load will be shown.
At this moment, the result of fine adjustment on
the minimum/maximum output of analog output
can be renewed.
7