126
1
Remove the connecting cable at the current output terminal.
2
Measure the voltage between the current output te and
−
.
Set the connecting range to DC
・
mA range for the measuring instrument such as
tester and so on.
3
Apply operational check on current output in the Check mode.
(Refer to the paragraph 7
−
18.)
Load resistance is
260 k
Ω
or less.
1
Put the connecting cable for the current output terminal where it was.
2
Check the setting related with current output.
●
F
−
20 Selection of target for the analog output.
●
F
−
21 Display value at the time of the minimum analog output.
●
F
−
22 Display value at the time of the maximum analog output.
(Refer to the paragraph 7
−
15)
Contact with Minebea.
Load resistance should be 260 k
Ω
or less.
Set correctly. Refer to the
paragraph 7
−
15, 8 and 9
−
1.
Set the HOLD input open.
Inform Minebea about the contents of
failure and situation at site in details.
10
NO
YES
Varies to 4 mA, 12
mA and 20 mA.
NO
YES
Setting is correct.
NO
YES
External control
input or HOLD is applied.
NO
YES