8
4
−
2. Note on connection
Warning
●
In case of connection with the instrument, keep strictly to the
following items. If neglected, it may cause an unexpected failure
or a damage to the instrument.
D
Be sure to set the power supply to OFF, when the connection will be made.
D
Since the terminal boards at rear side of the instrument is made of resin, take care not to
drop it down or not to apply strong impact.
D
Recommended torque to tighten the terminal screws for terminal board should be as
follows
:
Torque to tighten
the terminal screws
Terminal board
0.6 N
・
m
D
The suitable crimp type terminal lugs for the terminal board are as follows:
Width of crimp
type terminal lugs
Suitable crimp type
terminal lugs
Terminal board
6.2 mm or less
1.25
−
3 or Y type 1.25
−
3.5
D
Connecting cable with the instrument should be away from the noise source such as power
supply line and/or I/O line for control and so on as far as possible.
D
Conduit wiring should be the type of exclusive one, and avoid using with another line
together.
D
All of the connections should be executed securely by referring to the Instruction manual
for the instrument.