2. Calibration procedures to register the output of strain gage applied transducer at the
time of zero and the maximum display
(1) Procedure by key operation
Warning
●
Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor may cause malfunction in the instrument and there may
exist damage in peripheral equipments.
Besides, even though calibration has been made, there may occur the
similar case when the result is not correct, so make calibration again.
●
During the calibration is executing, be sure to set Tare weight
cancellation clear, and to make cancellation (Execution of F
−
98) for
compensated data on zero set and set the OFF position of Zero
tracking(Setting “00000” on F
−
07 and F
−
08), and also set the OFF
position of Peak.
If neglected, correct measured results may not be obtained.
●
During calibration procedures, press the
key in case of
interrupting the calibration is required. The calibration data will be
kept as they are before entering the calibration and then returns to
the Measurement mode.
●
Every time the
key is pressed with the load display of “FUNC”,
the display will change as the following arrow marks. However, every
time the
is pressed, the display will change as the reverse
direction of the following arrow marks.
“FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”
→
“SPAN”
→
“TARE”
→
“CHEK”
→
“MONT”
→
“VCAL”
→
“VADJ”
→
“FUNC”
→
“CCAL”
→・・・・・
(Hereinafter, it will repeat.)