7
−
2. Comparator
The instrument provides with the comparator by two set values of S1 and S2.
●
The comparator of the instrument executes the comparison operation
synchronizing with the display.
7
−
2
−
1. ON/OFF of comparator S1 and S2
Operational selection of ON/OFF can be made for each comparator S1 and S2. These selections
are done by the function (related function F
−
30). Default has selected both S1 and S2 turning
on.
7
−
2
−
2. Change of set value
Warning
●
When the set value for the comparator is set wrong, or set in the
wrong procedures, it may not obtain the correct results from the
comparator, and it may cause malfunctions in peripheral equipments
and also cause a damage as well.
Procedures
1
Change of S1 set value
In the Measurement mode, pressing the
key
makes the 10
0
digit and the judgement display S1
on the front panel load display will flash on and
off, then change of S1 setting can be available.
Every time the
key is pressed, the flashing
digit will change from 10
0
digit
→
10
1
digit
→
10
2
digit
→
10
3
digit
→
10
4
digit
→
Polarity(POL)
→
10
0
digit.
Change the set value by the
key after
flashing the digit which is desired to change.
If you keep the instrument untouched for
approx. 20 seconds in the condition of change
of setting is available, the Measurement mode
can be re
−
entered automatically.
After fixing up the change, press the
key.
Make the set value effective, then the
Measurement mode can be entered again.
:
Set value carry key
:
Set value inclement key
:
Set value initialization key
(Make the set value to “0”.)